Tecan Hybridization Station HS 400 Pro

$2,195.00

Streamline microarray processing with the Tecan HS 400 Pro Hybridization Station. Fully automated for reliability and precision. Available at BaneBio!

1 in stock

SKU: 240376 Category:

Description

  • System provides automated processing of microarrays on slides, from pre-hybridization and on-board denaturation, to automatic slide drying
  • The Tecan HS 400 PRO has been designed to avoid hands-on operation and give researchers better reproducibility, sensitivity and reliability than they can expect from traditional procedures with even less labor involved. It offer standardization and strict control of the entire process that are key factors in enhancing data quality and consistency. Its performance maximizes the value of genomic and proteomic research, increasing productivity in the laboratory. Low chamber volumes also help to save precious samples and cost
  • Temperature control: incubation/wash temperature range 4–85 °C (±1.0 °C)
  • denaturation temperature range 50–99 °C; max. 10 min
  • temperature precision ±0.5 °C
  • heat/cool rate >6 °C/min
  • precipitation prevention four reagent bottles pre-heated at 35–40 °C
  • Chambers
  • hybridization area  single area:  QuadChamber: 22 x 13 mm volumes QuadChamber: 55 µl
  • flow rate (washing) 11 ml/min (per slide)
  • liquid agitation six different modes, plus no agitation (including viscous buffers)
  • on slide nitrogen drying 2.7 bar N2
  • Slides per run
  • HS 400 Pro 1–4 slides; multiple instruments connected to one PC
  • Reagent bottles volume 6 x 1.0 liter
  • Dimensions
  • HS 400 Pro 280 mm x 430 mm x 550 mm (H x W x D)
  • Electrical power supply 100–240 V; 50/60 Hz (auto sensing) interface to PC serial RS 232; USB V1.x
  • Power consumption HS 400 Pro 300 VA

Additional information

Weight 50 lbs
Dimensions 25 × 25 × 25 in