Description
- The Biorad MyCycler Thermal Cycler is a sleek 96-well PCR system with a gradient feature and is used for gene amplification. This compact thermal cycler features intuitive programming of temperatures, times, and protocols with ready-made templates for ease of viewing, editing, and running protocols. The MyCycler’s self-adjusting heated lid maintains accurate temperature and accommodates standard 0.2 ml thin-wall PCR tubes, strips, and plates.
- Temp range: 4-100C
- Overshoot – Less than0.5°C
- Accuracy +/-0.5°C to NIST traceable standard after 30 seconds at 90°C, to block mode
- Uniformity -Well to well uniformity of +/- 0.5°C within 30 seconds after arrival at 90°C
- Input Gradient Range 1-25°C
- Functional gradient range- 5-25°C
- Minimum temperature of gradient-30°C
- Maximum temperature of gradient: 99°C
- Gradient accuracy: +/-0.5°C
- Gradient Uniformity: +/-0.5°C
- 100-120/220-240V 50-60Hz 910VA