Description
- Equipment connects and communicates with the software
- Buyer is responsible for obtaining computer and software
- Detection modes: UV-Vis
- Read methods: Endpoint, kinetic, spectral scanning
- Microplate types: 96- and 384-well plates
- Temperature control: 4-Zone incubation to 50 °C +/- 0.2 °C at 37 °C
- Shaking: Linear
- Software: Gen5 Data Analysis Software
- Absorbance
- Light source: Xenon flash
- Detector: photodiode
- Wavelength selection: monochromator
- Wavelength range: 200 to 999 nm, in 1 nm increments
- Monochromator bandwidth: 5 nm
- Dynamic range: 0 – 4.0 OD
- Resolution: 0.001 OD
- Pathlength correction: yes
- Monochromator wavelength accuracy: +2 nm
- Monochromator wavelength repeatability: +0.2 nm
- OD accuracy: 1% + 0.01 OD
- OD linearity: +1%
- Reading speed (kinetic): 96 wells: 5 seconds, 384 wells: 11 seconds
- Power: 100-240 Volts AC. 50/60 Hz.
- Dimensions: 8.5″W x 16″D x 8.5″H (21.6 x 40.6 x 21.6 cm)
- Weight: 24 lbs (11 kg)
- Environment: Operational temperature 18°-40°C
- Humidity: 10% to 80%, non-condensing
- Incubation option: Temperature Control: 4°C over ambient to 50°C
- Temperature Variation: ± 0.5°C across the plate @ 37°C
- Communication: One serial COM port (25-pin male connector)