Caron Photostability Chamber 6545-2

$14,995.00

1 in stock

SKU: 240417 Category:

Description

The 6540 Series Photostability Chambers were specifically designed to perform near UV and visual
light testing with fluorescent lamps. This second generation
bench-top unit controls light and temperature conditions through an easy-to-use color
touch-screen interface. Lights automatically shut off after a specified exposure level or
time duration. Uniform light distribution and high intensity levels allow quick response
for forced degradation testing and confirmatory studies. Precise humidity control
(Models 6545-1, 6545-2 and 6545-3) is achieved through ultrasonic injection and
mechanical refrigeration methods.

  • Independent control of near UV and cool white VIS lamps are
    provided to eliminate overexposure for confirmatory tests
    and provide flexibility for forced degradation studies. 
  • In “Automatic” mode, lights turn off at the end of a programmed
    exposure or timed test. 
  • “Manual” mode allows a flexible test time duration and makes
    validating the chamber easy. 
  • Dimmable lamps (230V models) allow testing at specific light
    intensity levels. 
  • Programmable alarms notify the user when aging lamps
    should be replaced. 
  • Specular aluminum reflective surfaces provide uniform UVA & VIS
    light reflection and maintain proper SPD (spectral power distribution). 
  • Stainless steel interior construction provides a smooth, non-porous
    surface for easy maintenance. 
  • Two light detectors (one VIS and one UVA) with an integral radiometer are
    standard and provide precise and accurate light intensity measurements.
    Light detectors are both cosine corrected and calibrated to NIST traceable standards. 
  • Precise temperature control is provided through use of a PID controller. 
  • Humidity controlled units (Models 6545-1, 6545-2 & 6545-3) isolate the effects of
    hydration during photostability testing.
  •  Shelves slide-out for quick and easy loading and unloading of samples. 
  • Strategic lamp spacing provides enhanced light uniformity. Overhead
    lighting results in a large illumination area for maximized direct lighting. 
  • Door safety switch turns lights off when the door is opened. The lockable
    door prevents accidental disturbance of in-process tests. 
  • Two access ports (one per shelf) are standard.
  • Lamp Type, UVA: Fluorescent, naer UV (Black Lamp)
  • Lamp Type, VIS: Fluorescent, cool white (iso 10977 compliant)
  • Light detectors: UVA and Visual, calibrated
  • UVA Light Intensity, Typical: 2-30 W/m2
  • VIS Light Intensity, Typical:  2-35 klux
  • Lamps Dimming: 3-100%
  • Light Uniformity:  UVA: ±10% VIS: ±10%
  • Temp Range: 10C to 35C (Lights on)
  • Temp Control: ± 0.2°C
  • Temp Uniformity: ± 2.5°C (lights on)
  • Humidity Range: 40-70% RH
  • Humidity Control: ± 3%
  •  Two (2) Sliding Shelves, 26.0″ W x 23.1″ D x 8″ H
  • Usable Shelf Area, Each:  19.0″ x 23.0″
  • Usable Shelf Area, Total: 6.1 ft2 (0.56m2
    ) – 76% of total shelf
  • Max weight per shelf: 22lbs
  • Internal Volume: 10.5 Cu Ft / 296L
  • Exterior Dims: 37.6″ W x 30.2″ D x 45.0″ H (955mm W x 767 mm D x 1133 mm H) w/leveling feet
  • Interior Dims:  26.9″ W x 24.0″ D x 28.0″ H (683mm W x 610mm D x 711mm H)
  • 208/230V 60Hz 15A