  • Measurement range: 10ug to 200mg of absolute water content
  • Titration speed: maxes out at 2.2 mg of H2O per minute
  • Generator cell configurations: available with or without a chemical diaphragm
  • Without diaphragm: highly convenient and virtually maintenance free for general organic solvents
  • With diaphragm: maximum precision for ultra low trace moisture (<100 ppm) or matrices causing anode side reactions
  • Repeatability: ±0.3% at water levels above 10mg; ±0.5% at water levels around 1mg
  • Automated thermal headroom gas extraction using a dry air or nitrogen carrier gas loop extraction method
  • Temp range: variable 40C to 280C adjustable in 1C steps
  • Temp scanning/ramp functionality: exclusive pro software module that automatically ramps temps across an unknown sample matrix to determine its ideal moisture evaporation point without decomposition
  • 26 position sample carousel optimized for 5mL glass sample vials
  • Carrier gas flow control: integrated digital mass flow controller adjusting gas sweeps between 10 to 150mL/min
  • Built in automated solvent handling system that empties and refills the titration cell at the click of a button to prevent user contact with toxic reagents
  • Integrated with Mettler Toledo LabX Titration software (sold separately from manufacturer)
  • Dimensions: 24.5″W x 17.5″D x 22.5″H
  • 100-240VAC 50/60Hz